How To Use Tempa Dot Disposable Thermometers

Tempa-Dot logo

Clinical Thermometers
SERVICING GUIDE — HOW TO USE/READ

STERILE CLINICAL  DISPOSABLE THERMOMETERS

3M MA5532 Tempa-Dot Sterile Clinical

Reduces the Risk of Infection

Temperature-taking devices tin be carriers of infectious diseases as they are shared between patients and handled by caretakers. Even with the most stringent cleaning and handling protocols, information technology is difficult to insure a contaminate-free device for each use. Tempa•DOT™ solves this problem by providing a sterile, individually wrapped, disposable thermometer. Simply open, use and dispose of. In the long history of Tempa•DOT™, in that location has not been a documentable instance of a patient-induced infection. A reduced infection rate tin translate into improved patient outcomes, and lower costs.

  • REDUCES THE RISK OF INFECTION— Other temperature-taking devices tin can accumulate, carry, and spread contagions as they are routinely handled. By comparison, Tempa•DOT™ is a sterile instrument, individually wrapped for unmarried use. This about eliminates the possibility of contaminating transfer. There is a strong case for the utilise of Tempa•DOT™ with all patients. The prevention of just i HAI issue can pay for the unabridged Tempa•DOT™ program. •
  • Authentic— Oral readings in ane infinitesimal. Accurate to ±0.1°C (within the critical range of 37.0°C to 39.0°C). • SAFE— Tempa•DOT™ is sterile, latex-free, and not-toxic. Individually wrapped to prevent contamination before use. •
  • CONVENIENT— Units are hands divisible and transportable. Place a 100-count box in each patient's room, or carry individual 4-strips. They fit comfortably in the pocket of your scrubs.
  • Economic— Eliminates time and materials spent to disinfect, recalibrate, recharge, or repair electronic devices. No demand to purchase and stock disposable (unsterilized) sheaths, or enter into long-term purchasing commitments.

HOW TO Store

Store Tempa•DOT™ thermometers in an area where temperatures do non exceed 30°C.
When exposed to high temperatures, matrix dots will turn blue. If this occurs:

  1. Place in freezer for 1 hour per box of 100, xvi hours per instance of 20 boxes. (2 hours per box for Tempa•DOT™ PLUS rectal thermometers.)
  2. Permit stand at room temperature for i day.
  3. Thermometers are now prepare for use. Dots may appear slightly off-color, merely this does non affect accuracy.

HOW TO OPTIMIZE RESULTS

As with all medical instruments, it is important to closely follow the IFUs:
– Position in mouth correctly.
– Continue your mouth closed for the full threescore 2d period.
– Read the concluding blue dot later on the device "locks in" (approx. 10 seconds). Ignore whatever skipped dots.3M MA5532 Tempa-Dot Sterile Clinical-OPTIMIZE

– Retake temperature with new device if left in the oral cavity longer than ii minutes (oral) or 5 minutes (axillary or rectal).
– Wait at least 15 minutes before taking temperature when patient is exposed to cold conditions, smoking, eating, or drinking.
– Do not utilise after the expiration date, which is printed on the 100 count box.
– If the thermometer wrapper has been damaged or opened, the product may no longer be sterile. Please discard.

ORAL Utilize

  1. Open PACKAGE
    Fold along perforation and tear off a unmarried Tempa•DOT™ thermometer strip. Peel back pinnacle of wrapper to expose handle of the thermometer. Remove thermometer by pulling the exposed handle, taking care not to affect the dot matrix portion which is to be placed in the patient's mouth.3M MA5532 Tempa-Dot Sterile Clinical-OPEN PACKAGE
  2. INSERT THERMOMETER AND Wait 60 SECONDS
    Place a thermometer into the patient'southward oral fissure under the tongue, as far back as possible into either estrus pocket, as you would whatsoever oral thermometer (see diagram). Generally, developed lips should bear upon the middle of the "Tempa•DOT" logo. The dots can face either up or down. Accept the patient press tongue down on the thermometer while keeping the mouth closed.
  3. REMOVE THERMOMETER ix 3 AND WAIT 10 SECONDS3M MA5532 Tempa-Dot Sterile Clinical-INSERT THERMOMETER
  4. READ AND Record TEMPERATURE
    The last blue dot on the matrix indicates the correct temperature (ignore any skipped dots). Retake temperature with the new device if left in the oral fissure for longer than 2 minutes.
  5. DISCARD THERMOMETER3M MA5532 Tempa-Dot Sterile Clinical-TEMPERATURE

Ordering Information:
Tempa•DOT # 5532 standard (°C Scale)
Tempa•DOT # 5538 standard (°C Scale)(UK only

AXILLARY Utilise

  1. Open Bundle
    Fold along perforation and tear off a single Tempa•DOT™ thermometer strip. Peel back peak of the wrapper to expose the handle of the thermometer. Remove the thermometer by pulling the exposed handle. 3M MA5532 Tempa-Dot Sterile Clinical-OPEN PACKAGE
  2. INSERT THERMOMETER AND WAIT 3 MINUTES
    Identify thermometer high in the armpit with the dot matrix side against the torso and parallel to the length of the body. Use the patient'due south arm to hold the thermometer in place.3M MA5532 Tempa-Dot Sterile Clinical-AIT 3 MINUTES
  3. REMOVE THERMOMETER AND Await 10 SECONDS
    Before reading, wait approximately 10 seconds for dots to lock-in. 3M MA5532 Tempa-Dot Sterile Clinical-10 SECONDS
  4. READ AND Tape TEMPERATURE
    The final blue dot on the matrix indicates the right temperature (ignore whatever skipped dots). Retake temperature with the new device if left underarm longer than 5 minutes.3M MA5532 Tempa-Dot Sterile Clinical-TEMPERATURE

Ordering Information:
Tempa•DOT # 5532 standard (°C Scale), or
Tempa•DOT # 5002 with an adhesive strip (°C Scale)
Tempa•DOT # 5538 standard (°C Scale)(the UK simply)

RECTAL Utilise

  1. Prepare THERMOMETER
    Inspect sheath. Practise not use if malformed or damaged. Insert thermometer completely into the sheath, dot-end showtime and with dots visible through the smooth side of the sheath. 3M MA5532 Tempa-Dot Sterile Clinical- PREPARE THERMOMETER
  2. INSERT THERMOMETER AND WAIT 3 MINUTES
    Lubricate the sheath exterior and insert the thermometer with the sheath into the rectum until the dots and numbers are completely covered.3M MA5532 Tempa-Dot Sterile Clinical-sheath exterior
    Caution: Whatsoever object inserted into the rectum could cause perforation. Exercise Not USE FORCE.
  3. REMOVE THERMOMETER AND Look 10 SECONDS
    Discard sheath. Before reading, look approximately 10 seconds for the dots to lock in.3M MA5532 Tempa-Dot Sterile Clinical-10 SECONDS
  4. READ AND Tape TEMPERATURE
    The concluding blue dot on the matrix indicates the correct temperature (ignore any skipped dots). Retake temperature with the new device if left in the rectum longer than v minutes.3M MA5532 Tempa-Dot Sterile Clinical-TEMPERATURE
  5. DISCARD THERMOMETER

Ordering Data:
Tempa•DOT # 5502R (°C Scale)

Documents / Resources

3M MA5532 Tempa-Dot Sterile Clinical Disposable Thermometers [pdf] User Guide
MA5532, Sterile Clinical Disposable Thermometers, MA5532 Sterile Clinical Disposable Thermometers, Clinical Dispensable Thermometers, Disposable Thermometers, Thermometers, MA5532 Tempa-Dot Sterile Clinical Dispensable Thermometers, Tempa-Dot Sterile Clinical Dispensable Thermometers

How To Use Tempa Dot Disposable Thermometers,

Source: https://manuals.plus/_3m/ma5532-tempa-dot-sterile-clinical-disposable-thermometers-manual

Posted by: emersonwaallovar.blogspot.com

0 Response to "How To Use Tempa Dot Disposable Thermometers"

Post a Comment

Iklan Atas Artikel

Iklan Tengah Artikel 1

Iklan Tengah Artikel 2

Iklan Bawah Artikel